Product Description:

TMS-2028 main technical parameter
◆ Volume (length× width ×height): 370 × 340 ×1570 mm ◆ UF flow scope: 0~1800ml/h(±30ml/h )
◆ Weight: about 90kg 0~4000ml/h(optional)
◆ Power supply voltage: •AC220V±10% ◆ ISO UF flow scope:0~2000ml/h(±30ml/h)
•Frequency: 50HZ ~ 60HZ ◆ TMP: •Scope:-100mmHg~+600mmHg
•Power: 1500W (±20mmHg)
◆ Blood pump/spare pump:•Flux: 15 ~ 340ml/min (Ф 6mm) ◆ Blood leakage monitor:
•20 ~ 460 ml/min (Ф 8mm) •over 1ml blood per liter dialysate
◆ Heparin pump: •Flux: 0.1ml/h~10ml/h(±5%) (flow:500ml/min)
•Precise: 0.1ml/h ◆ Blood level monitor: ultrasonic sensor
•Injector size: 20ml/30ml/50ml ◆ Air bubble monitor:
10ml/20ml/30ml(selectable) •Infrared and response threshold value:
◆ Arterial pressure: •display scope:-300 ~+300mmHg Single air bubble of 200 и I exists when
(±10mmHg) Blood flux is 200ml/min
◆ Venous pressure:•display scope:-50 ~ +300mmHg ◆ Inflow pressure:0MPa~0.6MPa
(±10mmHg) ◆ Inflow temperature: 5℃~30℃
◆ Dialysate flow:•300ml/min~800ml/min linearity ◆ Environment temperature: 10℃~30℃
adjustable (+10%) • Relative humidity≤70%
(-5%) ◆ Rinse/disinfection: chemical disinfection
•Temperature:35.0℃~39.0℃ (citric acid,peracetic acid and oxalic acid)
•Resolving rate:0.1℃ ◆ Hot rinse: 80℃
•Conductivity:13mS/cm~15.5mS/cm ◆ Back-up power supply: last for 15-30min
(±0.1 mS/cm) after electric-cut.(optional)
TMS-2028 function
Be applicable of hemodialysis treatment,hemoperfusion,Plasma Exchange,Pure ultrafiltration project.
Touch screen was equipped and it can show Na+and UF profiles;In addition,the functions ie.backup
power supply, Kt/V,HCO3¯ adjustment,single needle and acetate treatment are optional.